E4800 Series Recirculating Heater/Chillers Site Requirements

Electricity

Ensure that a suitable mains electricity supply (110 Vac or 240 Vac, frequency 50/60 Hz) is available. Check that the voltage label attached on the rear of the cabinet is suitable for the local voltage and frequency.

Site Selection

Take care when positioning the unit. Always leave open access to the front and rear of the chiller to allow adequate ventilation to the system. Ensure that the back meshes are not blocked with dust. The internal electrical fan is used to draw air through the vent grills. The warm air is expelled from the unit, which will raise the temperature of a small room. Try to site the unit so as to exhaust this air from the room, or provide adequate ventilation in the room.

Pumps

It is highly recommended that units fitted with high-pressure pumps should be fitted with a 50-micron water filter attached to the inlet connection of the unit.
If the system is fitted with a high-pressure pump, the connecting pipework and fittings must be able to take up to 5 bar (75 psi).

Flow Switches

For units ordered with optional Water Failure Alarms. The flow switches can be wired into the customer’s primary equipment so that, in the absence of fluid flow, the customers unit will also cut out until flow is resumed. A push button momentary switch is provided on the right hand side of the front panel to
by-pass the flow switch in order to initially start the system. The flow switch must be fitted on manufacture. The flow switch contacts require a 10 amp supply in the customers primary unit.

Space requirement

E4860: 320mm wide x 460mm deep x 370mm high Weight: 40Kg
E4870: 380mm wide x 610mm deep x 450mm high Weight: 62Kg
E4880: 450mm wide x 620mm deep x 500mm high Weight: 70Kg
E4890: 450mm wide x 620mm deep x 500mm high Weight: 82Kg

 